FKT 1DP1A-SV Advanced Pressure, Velocity and Flow Meter
|
The FKT 1DP1A-SV is portable, self-contained and micro-processor based. This instrument allows for the simultaneous measurement and display of differential pressure, atmospheric pressure, target gas temperature, relative humidity (RH), humidity ratio and density. Density is determined from the temperature and humidity readings. Accurate measurements can also be done on gases other than air by the user modifying the molecular weight of the target gas. In conjunction with a Pitot Static Tube, the FKT 1DP1A-SV can display measured velocity accounting for true gas density.
The FKT 1DP1A-SV can be used to perform duct surveys with built-in integration of the velocity profiles. In addition, this model estimates the uncertainty in your measurements for differential pressure, velocity and volumetric flow rate. The instrument also interfaces with a PC using the FlowScan software suite allowing data logging and flow survey functions and can configure as a component of a data acquisition system.
Features
- Simultaneous measurement and display of
- Differential pressure or velocity.
- Temperature.
- Relative Humidity.
- Humidity ratio.
- Atmospheric pressure (or absolute pressure).
- Real time display of true density including effects of temperature and water vapor.
- Excellent accuracy.
- Duct surveys at the touch of a button. Measure volumetric and mass flow rates in 3 user selectable units. SCFM available for ft/m range selection.
- Estimation of measurement uncertainty.
- Extreme ease of use. The instrument was specifically designed to avoid menu driven operation, as this usually proves cumbersome and requires continual reference to the manual. Almost all required information is displayed concurrently through the use of a large 4 line LCD display.
- K factor for non-unity probes. Allows use of S type probes for dirty environments.
- Pressure displayed in four user selectable units.
- Velocity is displayed in three user selectable units and with temperature and humidity corrections. In conjunction with a Pitot Static tube velocity based on true gas density can be displayed.
- Variable damping for measurements involving pulsating flows.
- Automatic auto-zeroing without disconnecting from the pressure source (through the use of internal solenoids) every ten minutes reduces zero-drift errors and simplifies operation. The auto-zero function can also be implemented by depressing the zero button. This is particularly useful when measuring in conditions with large ambient temperature fluctuations.
- Over-pressure protection: If the applied pressure exceeds the transducer rating, the instrument will automatically enter the auto-zero cycle which will equalize pressure across the differential transducers reducing the possibility of damaging the transducer.
- Data logging and flow surveying with optional FlowScan software suite.
- The absolute pressure transducer, if left unconnected, displays local atmospheric pressure, or if connected to a duct system the duct absolute pressure, which is then used for density calculation.
- The detachable probe of the instrument allows for the measurement of test point temperature and humidity.
- Rugged water resistant case.
Specifications
Enclosure |
UV protected, water resistant and virtually unbreakable |
Dimensions |
Length: 8.7 in (220.9 mm)
Width: 7.5 in (190.5 mm) Height: 3.9 in ( 99.1 mm) |
Weight |
2.93 lb (1.33 kg) |
Working Temperatures |
Operating: 32°F to 158°F (0°C to 60°C)
Storage: 14°F to 140°F (-10°C to 60°C) |
Power Supply |
8 x 1.5V AA Alkaline batteries, field replaceable
(batteries not included)
100V-240V AC auto-switching power supply (included) |
Battery life |
30hrs approx. |
Pressure Connector |
1/8 in barb, 0.41 in (10.45 mm) long by 0.19 in (4.9 mm) diameter.
Accepts 1/8 in ID rubber tubing |
Differential pressure |
For available pressure ranges see ordering information
One independent differential pressure transducer installed on each instrument Zero offset and zero temperature shift: eliminated through auto-zero Accuracy at 25°C: Typically within ±0.1% of Full Scale (±0.22% max). Media: Clean, dry, non-corrosive gases User recalibration procedure available |
Velocity |
For available velocity ranges see ordering information
Corrected for gas density and humidity Using a Velocity probe with user selectable flow coefficient User selectable gas molecular weight |
Absolute pressure |
Range: 2.2 -16.7 psi (15 – 115 kPa)
Accuracy (0 to 85°C): ±0.5% of Full Scale typical, includes effects of linearity, temperature and pressure hysteresis, zero temperature shift and span temperature shift. Media: Clean, dry, non-corrosive gases |
Damping |
User selectable from 1 to 64 data averages available. |
Temperature
|
K-Type flexible thermocouple with a range of -100°F to 900°F (-73°C to 482°C) The instrument can work with other K-type thermocouples and can display temperatures from -328°F to 1430°F (-200°C to 777°C) Accuracy: ±1.8°F (±1°C) Quick disconnect miniature size connector Cable length: 60 in (150 cm) |
Relative Humidity(detachable probe) |
Range: 0% to 99% Relative Humidity, non-condensing
Accuracy: ±2% typical at 25°C, non-condensing Storage: 0 to 90% Relative Humidity, non-condensing Response time (1/e): 15s in slow moving air at 25°C Cable length: 60 in (150 cm) Working temperature: -40°F to 185°F (-40°C to 85°C) Probe size: Cylinder 4 in (100 mm) long with a diameter of 0.5 in (13 mm). |
Display |
4-line large character variable contrast alphanumeric
LCD
Viewing area: 4.02 in (102 mm) by 1.63 in (41.5 mm) Pressure units: kPa, inH2O, mmHg and psi Velocity units: m/s, ft/s and ft/min Temperature units: °C and °F Density units: kg/m3 and lb/ft3 Humidity units: percent & mass ratio (kg of water vapor / kg dry air) Volumetric flux units: m/s, ft/s, ft/m, sft/m (scfm) Mass flux units: kg/m2s, lb/ft2s, lb/ft2m |
Output |
RS232 serial port interface, 9 pin connector. USB adapter available. |
Ordering Information & Accessories
All prices are given in USD and are subject to change without notice. Please contact us about applicable discounts.
Specifications can be customized. Please contact us for details.
Meter Part Number |
Description |
Price |
||||||||||||||||||||
FKT 1DP1A-SV-R Where R is the pressure range from the table below.
Order Example: FKT 1DP1A-SV-12 is an instrument with one ±12inH2O transducer. |
Full system containing
|
$1595.00 | ||||||||||||||||||||
Velocity values are for standard air conditions and limited to incompressible flow. See FKT 2DP1A-C for compressible high-speed flow capabilities. Other ranges available |
||||||||||||||||||||||
Accessory |
Description |
Price |
| FlowScan Software Suite | Software for logging and flow surveys | $295.00 |
| USBA | USB to serial adapter for computers with no serial port. Length: 1ft (30cm). Windows 98/2000/ME/XP. | $25.00 |
| CABLE6 | Serial cable replacement - 6ft (1.8m) | $10.00 |
| CABLE50 | 50ft (15.2m) serial cable. | $30.00 |
| Probes | Needed if you are measuring velocity or flow | Link |
| Filter Kit | Set of two 30-micron filters with replaceable elements. Includes a pack of 10 replacement filter elements. Useful for protecting the instrument in dirty environments. | $50.00 |
WPA |
Replacement auto-switching power supply (100V-240V AC 50-60Hz) US plug included |
$40.00 |
| Plug-EUR | European plug for power adapter | $7.00 |
| Plug-UK | UK plug for power adapter | $7.00 |
| Plug-AUST | Australian plug for power adapter. | $7.00 |