FKT 2DP1A-C High-Speed Meter

FKT 2DP1A-C

The FKT 2DP1A-C is portable, self-contained and micro-processor based. This meter allows for the simultaneous measurement and display of two independent differential pressures as well as concurrent measurement and display of atmospheric pressure, target gas temperature, relative humidity (RH) and density. Density is determined from the temperature and humidity readings. Accurate measurements can also be done on gases other than air by the user modifying the molecular weight of the target gas. In conjunction with a Pitot Static Tube, the FKT 2DP1A-C can measure velocity accounting for true gas density, high speed compressibility effects and user-selectable ratios of specific heats for gases other than air. The instrument also interfaces with a PC using the FlowScan software suite allowing data logging and flow survey functions and can be configured as a component of a data acquisition system.

Features

Specifications

Enclosure

UV protected, water resistant and virtually unbreakable

Dimensions

Length: 8.7 in (220.9 mm)

Width: 7.5 in (190.5 mm)

Height: 3.9 in ( 99.1 mm)

Weight

2.93 lb (1.33 kg)

Working Temperatures

Operating: 32°F to 158°F (0°C to 60°C)

Storage: 14°F to 140°F (-10°C to 60°C)

Power Supply

8 x 1.5V AA Alkaline batteries, field replaceable (batteries not included)

100V-240V AC auto-switching power supply (included)

Battery life

30hrs approx.

Pressure Connectors

1/8 in barb, 0.41 in (10.45 mm) long by 0.19 in (4.9 mm) diameter.

Accepts 1/8 in ID rubber tubing

Differential pressure

For available pressure ranges see ordering information

Up to two independent differential pressure transducers installed on each instrument

Zero offset and zero temperature shift: eliminated through auto-zero

Accuracy at 25°C: Typically within ±0.1% of Full Scale (±0.22% max).

Media: Clean, dry, non-corrosive gases

User recalibration procedure available

Velocity

For available pressure ranges see ordering information

Corrected for gas density, humidity and compressibility effects

Using a Velocity probe with user selectable flow coefficient

User selectable gas molecular weight

User selectable ratio of specific heats for the target gas

Absolute pressure

Range: 2.2 -16.7 psi (15 – 115 kPa).

Accuracy (0 to 85°C): ±0.5% of Full Scale typical, includes effects of linearity, temperature and pressure hysteresis, zero temperature shift and span temperature shift.

Media: Clean, dry, non-corrosive gases

Damping

User selectable from 1 to 64 data averages available.

Temperature
(detachable probe)

K-Type flexible thermocouple with a range of -100°F to 900°F (-73°C to 482°C)

The instrument can work with other K-type thermocouples and can display temperatures from -328°F to 1430°F (-200°C to 777°C)

Accuracy: ±1.8°F (±1°C)

Quick disconnect miniature size connector

Cable length: 60 in (150cm)

Relative Humidity
(detachable probe)

Range: 0% to 99% Relative Humidity, non-condensing

Accuracy: ±2% typical at 25°C, non-condensing

Storage: 0 to 90% Relative Humidity, non-condensing

Response time (1/e): 15s in slow moving air at 25°C

Cable length: 60 in (150 cm)

Working temperature: -40°F to 185°F (-40°C to 85°C)

Probe size: Cylinder 4 in (100 mm) long with a diameter of 0.5 in (13 mm).

Display

4-line large character variable contrast alphanumeric LCD

Viewing area: 4.02 in (102 mm) by 1.63 in (41.5 mm)

Pressure units: kPa, inH2O, mmHg and psi

Velocity units: m/s, ft/s and ft/min

Temperature units: °C and °F

Density units: kg/m3 and lb/ft3

Humidity units: percent

Output

RS232 serial port interface, 9 pin connector. USB adapter available.

Ordering Information & Accessories

All prices are given in USD and are subject to change without notice. Please contact us about applicable discounts.

Specifications can be customized. Please contact us for details.

Meter Part Number

Description

Price

FKT 2DP1A-C-R-R


Where R is the pressure range from the table below.

Order Example:

FKT 2DP1A-C-1-12 is an instrument with a ±1 and ±12inH2O transducers.



Full system which contains

  • Up to two differential pressure transducers (for ranges see table below)
  • Temperature sensor
  • Humidity sensor
  • Absolute transducer
  • Neck/shoulder strap
  • Auto-switching power supply with USA plug
  • Carrying bag
  • Printed manual
  • Tubing
  • NIST-traceable calibration with data.

$1795.00

two transducers


$1595.00

one transducer

R Differential Pressure / Velocity Range
0 No transducer
0.4 ±0.4inH2O (±100Pa) , 63 - 2536 ft/min , (0.3- 12.9 m/sec)
1 ±1.0inH2O (±249Pa) , 100 - 4000 ft/min , (0.5 - 20.3 m/sec)
5 ±5.0inH2O (±1.2kPa) , 224 - 8964 ft/min , (1.1 - 45.5 m/sec)
12 ±12inH2O (±3kPa) , 347 - 13887 ft/min , (1.7 - 70.5 m/sec)
20 -4 to +20inH2O , (-995 to +4.98kPa) , 448 - 17928 ft/min , (2.3 - 91.0 m/sec)
60 ±60inH2O (±15kPa) , 776 - 31052 ft/min , (4.0 - 157 m/sec)
100 -20 to +100inH2O , (-4.98 to +24.9kPa) , 1002 - 40089 ft/min , (5.0 - 203 m/sec)
180 ±180inH2O (±45kPa) , 1345 - 47252 ft/min , (6.8 - 240 m/sec)
360 ±360inH2O (±90kPa) , 1902 - 47252 ft/min , (9.7 - 240 m/sec)

Velocity values are for standard air conditions and limited to subsonic flow.


Other ranges available

Accessory

Description

Price

FlowScan Software Suite Software to read meter values into a computer $295.00
USBA USB to serial adapter for computers with no serial port. Length: 1ft (30cm). Windows 98/2000/ME/XP. $25.00
CABLE6 Serial cable replacement - 6ft (1.8m) $10.00
CABLE50 50ft (15.2m) serial cable. $30.00
Probes Needed if you are measuring velocity or flow Link
Filter Kit Set of two 30-micron filters with replaceable elements. Includes a pack of 10 replacement filter elements. Useful for protecting the instrument in dirty environments. $50.00

WPA

Replacement auto-switching power supply (100V-240V AC 50-60Hz)

US plug included

$40.00
Plug-EUR European plug for power adapter $7.00
Plug-UK UK plug for power adapter $7.00
Plug-AUST Australian plug for power adapter. $7.00